Generic Name: BioFire¿ Joint Infection (JI) Panel Generic Name: BioFire¿ Joint Infection (JI) Panel Model/Catalog Number: RFIT-ASY-0138 (30 Kit), RFIT-ASY-0135 (6 Kit) Software Version: N/A Product Description: Mulitple-type osteoarticular pathogen nucleic acid IVD, kit, nucleic acid technique (NAT) Component: N/A Reason: Due to potential contaiminated of s…
Medications
394 reports
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic Reason: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. Class IThe FDA's most serious recall level: a reasonable chance the product could cause serious health problems or death. · Ongoing · Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Swit…
Seen this too? Add your take
IntelePACS, InteleConnect Portal/Tech Portal Versions PACS-5-1-1-R102, PACS-5-2-1-R93, PACS-5-3-1-R83 through R128, PACS-5-4-1-R66 through R101, PACS-5-5-1-R50 through R89, PACS-5-6-1-R41 through R81, PACS-5-7-1-R21 through R61, PACS-5-8-1-R2 through R42, PACS-5-9-1-R8 through R19; System, Image Processing, Radiological Reason: Following a Chromium-based we…
Seen this too? Add your take
TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · These trays…
Seen this too? Add your take
Medline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K; 2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J; 3) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615L. Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dim…
Seen this too? Add your take
Convenience Kits containing CONMED AirSeal Access Ports KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G Reason: Kits contain CONMED AirSeal Access Ports. The ports were recalledThe maker or a government agency pulled this product. Check the model or lot number and stop using it until it's fixed or refunded. by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwid…
Seen this too? Add your take
The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. Reason: In vitro diagnostic analyzer had a risk of potentially inaccurate results r…
Seen this too? Add your take
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current o…
Seen this too? Add your take
iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm) Reason: The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution in the states…
Seen this too? Add your take
REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm Reason: Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Seen this too? Add your take
Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697C; 3) FEM POP PACK-LF, Medline Product Number/SKU DYNJ0159820F; 4) AV FISTULA PACK, Medline Product Number/SKU DYNJ67139D; 5) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AD; 6) LAMINE…
Seen this too? Add your take
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683 Reason: The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid…
Seen this too? Add your take
JenaValve Trilogy Delivery Catheter, Size 27MM REF: TRILOGY-DC-L Reason: Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Ger…
Seen this too? Add your take
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115 Reason: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if requir…
Seen this too? Add your take
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP Reason: Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) …
Seen this too? Add your take
Generic Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006 Product Description: The GORE VIABIL Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and se…
Seen this too? Add your take
JenaValve Trilogy Delivery Catheter, Size 25MM REF: TRILOGY-DC-M Reason: Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Ger…
Seen this too? Add your take
PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H Reason: Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced sut…
Seen this too? Add your take
Medline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR PACK-LF, Medline product number/SKU DYNJ0665837G; 3) AV FISTULA, Medline product number/SKU DYNJ900454C; 4) VEIN PACK SHD, Medline product number/SKU DYNJ68420D; 5) PACK VASCULAR BYPASS, Medline product number/SKU DY…
Seen this too? Add your take
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117 Reason: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if requir…
Seen this too? Add your take
FORWARD Gunshot Wound Kit; Model: AFT-GSW; Reason: Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the…
Seen this too? Add your take
Generic Name: FilmArray¿ Pneumonia Panel plus Generic Name: FilmArray¿ Pneumonia Panel plus Model/Catalog Number: RFIT-ASY-0143 (30 Kit), RFIT-ASY-0142 (6 Kit) Software Version: N/A Product Description: Multiple-type respiratory pathogen nucleic acid IVD, kit, nucleic acid technique (NAT) Component: N/A Reason: Due to potential contaiminated of spefic lots …
Seen this too? Add your take
1) Product Name AUD OPEN HEART PACK, Medline Kit Number/SKU DYNJ22052I; 2) Product Name AV SHUNT PACK-MARYVIEW, Medline Kit Number/SKU DYNJ24528N; 3) Product Name AV SHUNT PACK-MARYVIEW, Medline Kit Number/SKU DYNJ24528O; 4) Product Name AV SHUNT PACK-MARYVIEW, Medline Kit Number/SKU DYNJ24528P; 5) Product Name CATH LAB ENDO AAA, Medline Kit Number/…
Seen this too? Add your take
Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F Reason: Kits contain CONMED AirSeal Access Ports. The ports were recalledThe maker or a government agency pulled this product. Check the model or lot number and stop using it until it's fixed or refunded. by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending pro…
Seen this too? Add your take
Curaplex Capnography EtCO2/O2 Oral/Nasal Cannula Multipack (10 pk), Item Number 670272-KIT Reason: Affected kits were assembled with pediatric nasal cannula instead of the intended adult nasal cannula. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution in the states of IA, IL, IN, KY, MI, MN, MO, ND, OH, UT.
Seen this too? Add your take
Convenience Kits containing CONMED AirSeal Access Ports HERNANDEZ DAVINCI ACC BUNDLE M DYNJ911258 KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116I ROBOTIC CHEST DYNJ910997 DYNJ910997A RR-DAVINCI PROSTATE ACCESSORY DYNJ63392D UROLOGY LAPAROSCOPIC - 51638 DYNJ908720B UROLOGY ROBOT ADD ON PACK DYNJ85189 Reason: Kits contain CO…
Seen this too? Add your take
BD Phoenix Panel NMIC-312. Catalog Number: 449082. Reason: Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution.
Seen this too? Add your take
Vinyl Instrument Guards, labeled as: 1) VENTED INSTRUMENT GUARD 1/16 X 3/4H - WHITE, Model Number 09-1-011; 2) WHITE VENTED INSTRUMENT GUARD 1/16 X 3/4, Model Number 09-1-011; 3) VENTED INSTRUMENT GUARD 5/64 X 3/4 - BLUE, Model Number 09-1-012; 4) BLUE VENTED INSTRUMENT GUARD 5/64 X 3/4, Model Number 09-1-012; 5) VENTED INSTRUMENT GUARD 7/64 X 3/4 - GREEN, …
Seen this too? Add your take
Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO Reason: The implantable clip may experience difficulties deploying, which could prevent the reflector from functioning as intended during use. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US: AZ, CA, FL, IN, LA, N…
Seen this too? Add your take
Medline surgical convenience kits labeled as: CABG AUXILIARY PACK, Medline Product Number/SKU DYNJ44925D Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use. Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · US Nationwide distribution.