RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
Official notice from a government agency or its published feed.
Reason: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · Worldwide - US Nationwide distribution.
Hazard: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the…
View the FDA enforcement report ↗ ← All reports