Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X…
Official notice from a government agency or its published feed.
Reason: Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Class IIA mid-level FDA recall: the product may cause temporary or reversible health problems. · Ongoing · Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MD, MN, NC, NY, OH, PA, TX, VA, WV and the countries of Australia, Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, Denmark, Ecuador, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lithuania, Macao, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Nicaragua, Norway, Palestine, Panama, Philippines, Poland, Portugal, Romania, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
Hazard: Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
View the FDA enforcement report ↗ ← All reports